Histopathological findings of the human great saphenous vein treated with endoluminal radio frequency ablation

Authors

  • Harinder Singh Bedi Department of Cardiovascular Endovascular and Thoracic Surgery, Christian Medical College & Hospital, Ludhiana, Punjab
  • Nalini Calton Department of Pathology, Christian Medical College & Hospital, Ludhiana, Punjab
  • Kanwardeep Singh Kwatra Department of Pathology, Christian Medical College & Hospital, Ludhiana, Punjab
  • Vivek Tewarson Department of Cardiovascular Endovascular and Thoracic Surgery, Christian Medical College & Hospital, Ludhiana, Punjab

Keywords:

Great saphenous vein, Radio frequency ablation, Endovenous therapy, Varicose veins

Abstract

Background: There is not much data on the histopathologic effect of Radio Frequency Ablation (RFA) on the human vein. In order to better understand the exact mechanism and sites of action of the device we studied in detail the histological changes in the vessel wall of the Great Saphenous Vein (GSV) after it was subjected to RFA.  

Methods: We studied the effect of RFA on 5 pieces of the human GSV from 5 different patients. The harvested vein was subjected to radiofrequency ablation and then studied in detail by histopathology.

Results: On gross inspection immediately during the radiofrequency application, the vein shrunk lengthwise and in diameter with marked narrowing of the lumen and charring. On histopathology the vein showed a photo-disruption in the form of total loss of endothelium. There was a sub-endothelial hematoma with small haemorrhages in the muscle coat and thermal coagulation of smooth muscle. The thermal effect was pan-mural and extending up to the adventitia.  

Conclusion: Our findings have implications for a better understanding of the mechanism of the action of radiofrequency on the venous system and for further improvements in technique and hardware to improve efficacy and reduce the side effects during endovenous therapy for varicose veins.

Metrics

Metrics Loading ...

References

Bremer JVD, Moll FL. Historical overview of varicose vein surgery. Ann Vasc Surg. 2010;24(3):426-32.

Weiss RA. Comparison of endovenous radiofrequency versus 810 nm diode laser occlusion of large veins in an animal model. Dermatol Surg. 2002;28:56-61.

Proebstle TM, Lehr HA, Kargl A, Epsinola-Klein, Rother W, Bethge S, Knop J. Endovenous treatment of greater saphenous vein with 940 nm diode laser: thrombotic occlusion after endoluminal thermal damage by laser generated steam bubbles. J Vasc Surg. 2002;35:729-36.

Schmedt CG, Sroka R, Steckmeier S, Meissner OA, Babaryka G, Hunger K, et al. Investigation on radiofrequency and laser (980 nm) effects after endoluminal treatment of saphenous vein insufficiency in an ex vivo model. Eur J Vasc Endovasc Surg. 2006;32(3):318-25.

Downloads

Published

2016-12-10

How to Cite

Bedi, H. S., Calton, N., Kwatra, K. S., & Tewarson, V. (2016). Histopathological findings of the human great saphenous vein treated with endoluminal radio frequency ablation. International Surgery Journal, 1(1), 3–5. Retrieved from https://www.ijsurgery.com/index.php/isj/article/view/245

Issue

Section

Original Research Articles